Specifications
- The monitor has abundant functions that can be used for clinical monitoring with adult, pediatric and neonate.
- Users may select different parameter configuration according to different requirements.
- The monitor, power supplied by 100-240V~,50/60Hz, adopts 12.1” color TFT LCD displaying real-time date and waveform.
- It can synchronously display eight-channel waveform and full monitoring parameters equipped with an optional 48mm thermal recorder.
- The monitor can be connected to the central monitoring system via wire or wireless network to form a network monitoring system.
Functions
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
1)ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis
2)RESP Respiration rate(RR)
Respiration waveform
3)SpO2 Oxygen saturation(SpO2)
Pleth ys mogram (PLETH) waveform
Pulse rate(PR)
Bar graph
4)NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean pressure(MEAN)
5)TEMP T1, T2, TD
6)IBP(optional) CH1:SYS,DIA,MAP
CH2:SYS,DIA,MAP
IBP waveform
7)CO2 EtCO2
InsCO2:Inspired Minimum CO2
AwRR:Air Way Respiration
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc.
Features
- 12.1” TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakh, Polish, Czech).
- fan less design, quiet, energy-saving and clean, which reduces the possibility of cross-infection.
- All-round monitor for adult, pediatric and neonate.
- With standard interface, oxygen graph, trend graph, big character interface and view bed, convenient to observe.
- Finish all operations by keys and knobs.
- Maximum 8-channel waveform synchronous display.
- Display 7-lead ECG waveform on one screen, cascade ECG waveform display
- Adopt digital SpO2 technology, anti-motion and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
- Heart rate variability (HRV) analysis function
- NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
- Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
- Drug concentration calculation and titration table functions.
- One-touch printing of trend graph
- Connect to Central Monitoring System by 3G, Wi-Fi or wired mode.
- AC/DC, built-in rechargeable lithium battery achieve uninterrupted monitoring.
- Anti-high frequency surgical unit, defibrillation-proof (special leads are necessary).
- Data saving up to 72 hours in the internal memory
- Up to 150 minutes use with a battery removable and rechargeable
Performance
▲ECG
Lead Mode | 3-Lead Or 5-Lead |
Lead Selection | I, II, III, AVR, AVL, AVF, V |
Waveform | 5-Lead, Dual-Channel |
Gain | 2.5mm/MV, 5.0mm/MV, 10mm/MV, 20mm/MV, 40mm/MV |
Scan Speed | 12.5mm/S, 25 Mm/S, 50 Mm/S |
HR | Measurement And Alarm Range: 15~350 Bpm |
Accuracy | ±1 % Or ±1 Bpm, Whichever Is Greater |
Alarm Accuracy | ± 2 Bpm |
Resolution | 1 Bpm |
ST-Segment Monitoring | Measurement And Alarm Range: -2.0 MV ~ +2.0 MV |
Accuracy | -0.8 MV~+0.8 MV ±0.04 MV Or ±10%, Whichever Is Greater |
Other Range | Unspecified |
Arrhythmia Analysis | ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC |
Pacemaker | Yes |
▲RESP
Method | R-F(RA-LL) Impedance |
Respiration Rate | Measurement And Alarm Range: 0~150 Rpm |
Resolution | 1 Rpm |
Measurement Accuracy | ±2 Rpm |
Alarm Accuracy | ±3 Rpm |
Apnea Alarm | 10~40s |
Scan Speed | 6.25 Mm/S, 12.5 Mm/S, 25 Mm/S |
▲NIBP
Method | Oscillometer |
Mode | Manual/AUTO/STAT |
Measurement Interval In AUTO Mode | 1/2/3/4/5/10/15/30/60/90/120/240/480/960 Minutes |
Measurement Period In STAT Mode | 5 Minutes |
Measurement And Alarm Range | 10 ~ 270 MmHg |
Resolution | 1 MmHg |
Cuff Pressure Accuracy | ±3 MmHg |
Measurement Accuracy | Maximal Mean Deviation: ±5 MmHg |
aximal Standard Deviation | 8 MmHg |
Over-Pressure Protection | Adult Mode: 297±3 MmHg Pediatric Mode: 240±3 MmHg Neonatal Mode: 147±3 MmHg |
▲SpO2
Measurement and alarm range | 0 ~ 100% |
Resolution | 1% |
Measurement accuracy | 70%~100%: ±2% |
unspecified | 0%~69% |
▲PR
Measurement and alarm range | 30 ~ 250 bpm |
Measurement accuracy | ±2 bpm or ±2%, whichever is greater |
▲TEMP
Channel | Dual-Channel |
Measurement And Alarm Range | 0 ~ 50℃ |
Resolution | 0.1 ℃ |
Accuracy | ±0.1 ℃ |
▲EtCO2
Method | Side Stream Or Mainstream |
Measuring Range | 0~150mmHg |
Resolution | 0~69 MmHg, 0.1 MmHg 70~150 MmHg, 0.25 MmHg |
Accuracy | 0~40 Mm Hg ±2 Mm Hg 41~70 Mm Hg ±5% 71~100 Mm Hg ±8% 101~150 Mm Hg ±10% |
AwRR Range | 2~150 Rpm |
AwRR Accuracy | ±1BPM |
Apnea Alarm | Yes |
Accessories
1)Adult fingertip SpO2 probe (5-pin)
2)Adult NIBP cuff
3)NIBP extension tube
4)ECG lead cable
5)ECG electrode
6)Temperature probe
7)Power cord
8)User Manual
Physical characteristic
Dimension | 314 Mm(L) × 145 Mm(W) × 264 Mm(H) |
Weight | 3.9 Kg |
Reviews
There are no reviews yet.