Possible Applications:
Biochemistry tests on whole blood, serum, plasma, or urine samples.
- Semi-automated photometric system
- Effective temperature regulation system
- Flexible cuvette concept, interchangeable flow-through or
standard cuvette
- Teach-in capability for reagent applications via touchscreen
- Minimal sipping volume 250 μl
- Double-secured liquid control with infrared bubble detector
- Future-proof operating system can easily be upgraded
TECHNICAL SPECIFICATIONS
Type | Semi-automatic, single-beam filter photometer |
Light Source | LED with long lifetime |
Wavelength | 340 nm and 380 nm – 730 nm |
Wavelength Selection | Automatic via 9-position filter wheel : 6 standard interference filters: 340 nm, 405 nm, 492 nm, 546 nm, 578 nm and 623 nm 3 positions for optional filter of choice |
Photometric Range | 0 – 2.5 A |
Cuvette System | Micro flow cell: 32 µl, 10 mm light path interchangeable with normal standard cuvettes (macro or semi-micro, disposable or special optical glass) |
Temperature Control | Internal Peltier element, temperature variable, pre-adjusted to 25 °C, 30 °C and 37 °C
Equilibration time for aspirated reaction mixture to reach 37 °C from ambient temperature: 15 sec |
Aspiration System | Built-in peristaltic pump driven by stepper motor, programmable aspiration volume controlled by infrared light barrier |
Sipping Volume | Minimum 250 µl, typically 500 µl up to 2000 µl
Separate setting of aspirate volume and wash volume |
Operator Interface | Touchscreen for direct functions and alphanumerical inputs |
Data Presentation | Graphic display: white characters or symbols, blue backlight, dimension 240 * 128 dots |
Integrated Printer | Thermal printer, 24 characters per line |
Languages | English and French / German / Indonesian / Russian / Spanish |
Memory | General operating software can be updated by PC
Reagent open system with capacity for up to 231 programmable methods Import of data by touchscreen or PC Up to 50 non-linear calibration curves with max 20 sets of points can be stored |
Data Logging | Up to 1000 results can be saved in the memory |
Signal Port | Serial port for connection to an external printer and/or PC
RS232 connector for barcode reader for an easy management of sample identities |
Measurement Procedures | Absorbance
End Point with factor, standard or multiple standards, with or without reagent blank and/or sample blank Bichromatic end point Kinetics with standard, factor or multiple standards, with or without blank Fixed time with standard, factor or multiple standards, with or without reagent blank Turbidimetry Single, double and triple determinations Curve fitting for non-linear standard curves Free hemoglobin in combination with optional interference filters |
Quality control | Up to 50 methods can be controlled with two control sera Levey Jennings plot |
Measuring Time | Kinetic: variable from 3 – 19 deltas, time per delta 4 – 255 sec in 1 sec steps
Fixed time: variable from 0 – 1800 sec in 1 sec steps |
Delay Time | Programmable from 0 – 1800 sec in 1 sec steps |
Mains Supply | Range: 100 VAC up to 240 VAC at 50/60 Hz |
Dimensions | Length 33 cm x width 34 cm x height 18 cm |
Weight | 5.0 kg |