Hemodialysis Machine 4008S NG-جهاز غسيل الكلى

POA
Request Price Now

The 4008S Next Generation has been developed to provide safety and efficiency for all important treatment modalities within the scope of renal replacement therapies.

Warranty period: 1 years

Description

Specification

  • The 4008S Next Generation has been developed to provide safety and efficiency for all important treatment modalities within the scope of renal replacement therapies. This includes:
  • Bicarbonate or acetate dialysis
  • Single-needle or single-needle click-clack
  • Ultrafiltration and sodium profiles
  • Isolated ultrafiltration / Bergström treatment
  • The automated acquisition, monitoring and control of specific patient and machine parameters with the 4008S Next Generation provides the operator with continuous and precise assessment of the current treatment situation.

Key features of the 4008S Next Generation include:

  • Monitoring of dialysis dose by OCM® (Online clearance monitoring and plasma sodium measurement)
  • Treatment with ultrapure dialysis fluid (DIASAFE®plus filter)
  • Use of hygienic, dry bicarbonate concentrate (Bibag®)
  • Single-needle method
  • Connection for central concentrate supply

In addition, the following therapeutic tools are available:

  • Blood pressure monitor (BPM)
  • Easy data management with patient card (offline) or network integration (online)
TypeBody Fluids Processing Equipment
Brand NameFresenius
Place of OriginGermany
Instrument classificationClass I
Dimensions1370x500x650(HxWxD)
Weight86 KG
Water supply Water inlet pressure1.5 to 6.0 bar
Water supply Water inlet temperature5°C to 30°C; for “integrated hot rinse” 85°C to 95°C
Water supply Max. drain height1 m
Concentrate Supply pressure0 to 100 mbar; 1 m max. suction height
Concentrate Central supply1 central acid concentrate (optional) 0 to 500 mbar
Power supply100 to 240 V AC, 50 to 60 Hz
Current consumptionApprox. 9 A (at 230 V) and 15 A (at 110 V)
External connectionsNetwork interface for data exchange with Therapy Data Management System (optional); input/output for connection of external
auxiliary equipment; alarm input/output (e.g. for nurse call); diagnosis for in-house computer diagnosis
Battery18 V; 3 Ah (maintenance free)
Arterial pressure monitoring Display range– 300 mmHg to + 280 mmHg
Arterial pressure monitoring Accuracy±10 mmHg
Arterial pressure monitoring Resolution20 mmHg
Venous pressure monitoring Display range– 60 mmHg to + 520 mmHg
Venous pressure monitoring Accuracy±10 mmHg
Venous pressure monitoring Resolution20 mmHg
Transmembrane pressure monitoring Display range– 60 mmHg to + 520 mmHg
Transmembrane pressure monitoring Resolution20 mmHg
Arterial blood pump Blood flow range15 to 600 mL/min
Arterial blood pump Accuracy± 10%
Arterial blood pump Resolution5 mL/min
Arterial blood pump Single needle system (optional)With 2 blood pumps, internal pressure/pressure control with variable stroke volume (max. 50 mL/min)
Arterial blood pump Air bubble detectorUltrasonic transmission measurement on blood line, additional optical monitoring in venous clamp
Arterial blood pump Heparin pumpDelivery range: 0 to 10 mL/h Bolus function: max. 5 mL per bolus Syringe size: 20 mL
Dialysis fluid flow range Dialysis fluid flow range0 – 300 – 500 – 800 mL/min
Dialysis fluid flow range Adapted Flow*(selectable) Dialysate flow adapted to the effective blood flow
Dialysis fluid temperature Selectable35°C to 39°C
Dialysis fluid conductivity Range12.8 to 15.7 mS/cm (25°C)
Dialysis fluid conductivity Accuracy± 0.1 mS/cm
Dialysis fluid conductivity Resolution0.1 mS/cm
Dialysis fluid acid component Mixing ratioAdjustable, e.g. 1+44, 1+34
Dialysis fluid acid component Adjustment range125 to 150 mmol/L
Dialysis fluid bicarbonate component Default mixing ratio1 + 27.6 (others possible)
Dialysis fluid bicarbonate component Adjustment range– 8 to + 8 mmol/L
Accurate Clearance K± 6%
Bicarbonate dry concentratebibag
Dialysis fluid filter systemDIASAFEplus
Balancing accuracy± 0.1 % according to the total dialysate volume
Pressure holding testsCyclic
Ultrafiltration UF rate0 to 4000 mL/h (in steps of 1 mL/h)
Ultrafiltration Pump volume accuracy± 1%
Ultrafiltration Parameters displayedUF goal, UF time, UF rate, UF volume
Blood leak detector Sensitivity ≤ 0.5 mL blood/min (Hct = 25) at max. flow 800 mL/min
 Disinfection and cleaning programs**
Rinse Temperature/flow37 °C / 600 mL/min
Hot rinse (recirculation) Temperature/flow84 °C / 450 mL/min
Integrated hot rinse Temperature/flow84°C / 450 mL/min
Cleaning Sporotal100 (recirculation) Temperature/flow37 °C / 600 mL/min
Heat disinfection Diasteril/Citrosteril (recirculation) Temperature/flow84° C / 450 mL/min
Disinfection Puristeril 340/plus (recirculation) Temperature/flow37 °C / 600 mL/min
Reviews (0)

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.