Specifications
Three operating modes – Rate Mode, Time Mode and Body Weight Mode
- Multiple modes for operators to choose
Wide range of injection rates
- Applicable to both adults and children, in normal and emergency cases
Built-in double CPU
- Each CPU can serve as backup for the other, guaranteeing the stability of the system and the safety of patients even when one CPU
- breaks down
Powerful alarm system with both audible and visible alarm signs
- Embedded pressure sensor for occlusion detection
- Multiple alarm signs to alert operators
Extreme flexibility in the use of syringes
- Almost all brands of standard syringes can be applied and automatically recognized by the module to save your cost and time
Wireless linkage with infusion supervision system
- Greatly reduces workload and turnaround time and ensures the safety of patients
Technical Specifications
- Syringe Type 5, 10, 20, 30, 50 ml
- Operating Mode
1) Injection rate mode, rate range: 0.1 – 2000 ml/h (increment 0.1 ml/h)
2) Time mode, time range: 1 – 6000 min
3) Body weight mode, weight range: 0.1 – 300.0 kg; drug limit: 0.1 – 999.9 mg; volume range: 0.1 – 999.9 ml
Accuracy | ± 5% |
KVO Rate | 1 – 5 ml/h |
Injection Rate Range 5 ml syringe | 0.1 ml/h – 100ml/h; 10 ml syringe: 0.1 ml/h – 200ml/h; 20 ml syringe: 0.1 ml/h – 400ml/h |
30 ml syringe | 0.1 ml/h – 600ml/h; 50 ml syringe: 0.1 ml/h – 1500ml/h |
Accumulated Injection Volume | 0.1 – 9999.9 ml |
Injection | ± 5% |
KVO Rate On/Off; On | 0.1 – 5.0 ml/h |
Bolus Rate 5 ml syringe | 100 ml/h; 10 ml syringe: 200 ml/h; 20 ml syringe: 400 ml/h; 30 ml syringe: 400 – 600 ml/h |
50 ml syringe | 400 – 1500 ml/h |
Display
- Screen 3.5-inch B/W LCD
- Information Mode, Rate, Accumulated injection volume, Syringe size, Battery capacity, Bed No., AC power indicator,
- Battery charge indicator
Alarm Type
- OVER, NEAR, Occlusion, Syringe disengaged, Pump stuck, No AC power supply, Low battery, Serious low battery, Abnormal 1/2/3
Battery
- Type 7.4 V, 1600 mAh lithium polymer battery
- Life More than 2 hours’ continuous running (5 ml/h); Rechargeable
Environment
- Temperature 5 – 40 ℃ for operating; -20 – 50 ℃ for storage
- Air Pressure 86 – 106 kPa for operating; 50 – 106 kPa for storage
- Humidity 20 % – 80 % for operating; 10 % – 95 % for storage
Others
Dimensione | 310 mm(L) × 125 mm(D) × 115 mm(H) |
Weight | < 3 kg |
Power Supply | 100 – 240 VAC, 50/60 Hz |
Shell Material ABS plastic
Software Language English, Russian, Portuguese
Reviews
There are no reviews yet.