Description

Friendliness

  • Compact and lightweight design for easy carrying and mounting
  • Long runtime battery, suitable for emergency and patient transfer
  • Intuitive operation and easy to install IV sets
  • IP34 protection

Safety

  • High accuracy:  ±5%
  • Dynamic pressure detection, Anti-bolus
  • High sensitivity air bubble detection: minimum 50l

Connectivity

  • Ambulance support
  • WiFi connection available
  • Dock support
Specifications and Functions
Dimension(L x W x H)150 x 90 x 200mm
Weight≤1.5kg
ClassificationCF, Class I
Screen2.5″ LCD monochrome screen, 132 x 32 pixels
Infusion modeRate mode
Flow rate range0.1~1500 ml/h
VTBI0.1~9999 ml
Accumulated volume0.1~9999 ml, increment is 0.1 ml
Increment0.1 ml/h
Accuracy≤ ± 5%
KVO rate0.5 ml/h, unadjustable
Bolus rate0.1-1500ml/h
Purge rate600ml/h, unadjustable
Occlusion150/525/900 mmHg
Air bubble detection5 levels selectable: 50µl, 100µl, 250µl, 500µl, 800µl
TitrationYes
Keypad lockYes
PauseYes
Anti-bolusYes
Self-test systemYes
Alarm Volume1-8 levels selectable
Auto lock“ON” and “OFF”, 1-5 min selectable, step 1 min, default “OFF”
Automatic recognition of IV setsYes
Mounting typeStandard: Vertical Pole; Optional: Horizontal Pole
Alarm
 Occlusion, VTBI Done, Battery Empty, VTBI Near Done, Reminder, Battery Low, AC Power Disconnection, System Error, System abnormal, KVO Finish, Standby Time Expired,

Tube not inserted, Air in line, Door opened, Empty bottle, Drop error and No communication

Power supply and Safety Requirement
AC power100-240V (50/60Hz)
Battery typeLithium
Battery operation timeStandard(1 battery):≥4 hours@25ml/h; Optional(2 batteries): ≥8 hours@25ml/h
Charging timeStandard(1 battery): ≤ 6hours; Optional(2 batteries):≤ 12hours
IP GradeIP 34
Temperature0~40ºC for operating, -40~70ºC for storage
Humidity15~95% for operating, 10~95% for storage
Air pressure57~106kPa for operating, 50~106kPa for storage
Connectivity
Data interfaceSupported, RS232
WirelessWiFi (optional)
Nurse CallSupported
DC inputSupported
AmbulanceSupported, meet EN-1789 requirement
Reviews (0)

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.